Spacelabs公司动态血压监测仪的精确性得到了验证

【 2005-07-22 发布 】 美迪医讯
美迪网领先的医疗器械电子商务平台

 
在持续应用了6年之后,以项研究证实了Spacelabs医疗公司的动态血压监测器(ABP)的精确性。

研究人员使用了模拟器来常规检查15台Spacelabs公司ABP设备的精确性(大多数型号是90207),这些设备是从1998~2004年在英国爱丁堡的西部总医院心血管风险诊所中使用的。研究方案设计用于测试设备在低血压、正常血压和高血压条件下的状态。检查记录的数据评估设备的远期的协调性和可靠性。

在被检测的1920套设备中,只有1套超出了Spacelabs使用模拟波形记录所预期的通常认可的压力范围,而且这套设备超出界限只有1mmHg。协调性标准要求设备记录的收缩压、舒张压、和平均动脉压的标准差在2mmHg以内,至少2/3在2mmHg以内。在测试中,设备记录到的平均压力有几乎90%的数值,在连续检测间隙都在2mmHg之内,绝大多数都小于1mmHg。

心血管风险诊所的内科医生Sister Gough解释说:“在西部总医院,在第一次临床咨询之前,每1位在心血管风险诊所的患者都接受了24小时动态血压监测。Spacelabs的动态血压监测在我们诊所得到了很好使用。监测器被计划在24小时的时间内,按照30分钟的循环记录血压,而且鼓励患者在日常活动中继续佩带监测器监测血压。我们的监测成功率超过了95%。”

西部总医院的维护记录显示,大多数血压监测设备需要中断维持的机会,每3年不到1次,没有设备需要维修的次数超过每年以一次。

Valdidating Blood Pressure Monitors
 
Following six years of continuous use, a study has validated the accuracy of the ambulatory blood pressure (ABP) monitors of Spacelabs Medical (Issaquah, WA, USA).

The researchers used a simulator to routinely check the accuracy of 15 Spacelabs ABP devices (most were model 90207) in use at the Cardiovascular Risk Clinic at the Western General Hospital in Edinburgh (UK) from 1998-2004. The protocol was designed to test the devices over a range of low, normal, and high pressures. The records were examined to assess the long-term consistency and reliability of the devices.

Out of the 1,920 sets measured, only one was outside the generally accepted limits of the pressure expected from the Spacelabs records with the simulated waveforms, and that one exceeded the limits by only 1 mmHg. The criteria for consistency call for the standard deviation of the systolic, diastolic, and mean arterial pressures recorded to be within 3 mmHg, with at least two-thirds within 2 mmHg. In the test, nearly 90% of the average pressures recorded by any device remained within 2 mmHg of their recordings at successive test intervals. Most differences were less than 1 mmHg.

“At the Western General Hospital each patient referred to the Cardiovascular Risk Clinic has 24-hour ambulatory blood pressure monitoring prior to their first clinical consultation,” explained Sister Gough, a physician at the Cardiovascular Risk Clinic. “Spacelabs ambulatory blood pressure monitors are used exclusively at our clinic. The monitors are programmed to record blood pressure measurements at 30-minute cycles over a 24-hour period and patients are encouraged to continue with their normal daily activities while wearing the monitor. We achieve a monitoring success rate in excess of 95%.”

Maintenance records at the Western General Hospital showed that most blood pressure monitoring devices required breakdown maintenance less than once every three years, with no devices requiring repair more than once a year.

/**/
收藏本文到: Google书签 Digg Live Bookmark Yahoo书签 Facebook 百度搜藏 新浪ViVi 365Key网摘 天极网摘 和讯网摘 POCO网摘 QQ书签

  《美迪医讯》欢迎您参与新闻投稿,业务咨询: 美迪医疗网业务咨询

网友评论
共计 1 条      
  网友 美迪网友 于 2005-08-13 21:32:59 发表评论 IP:60.171.111.224 信箱
    我想买一台!那有卖?
 
我要评论:《Spacelabs公司动态血压监测仪的精确性得到了验证》
匿名发表 我的名字: Email: 验证码: 点击可刷新
 
    
合作支持:中华医学会 | 中华医院管理学会 | 国家食品药品监督管理家用护理器械商城 | 国药励展展览有限责任公 | 医学装备协会
刊登广告 | 友情链接 | 广告代理商加盟 | 关于美迪 | 法律声明 | 隐私保护 | 网站地图
把美迪网放进收藏夹  把美迪医疗网介绍给我的朋友  给美迪医疗网留言
美迪医疗网广告业务联系:021-51601230 产品咨询业务联系:021-51601230 传真:021-56532303    美迪医疗网业务咨询
互联网药品信息服务许可证:(沪)-经营性-2009-0003   中华人民共和国电信与信息服务业务经营许可证:(沪)B2-20090029 沪ICP备14001091号-8
 
公安备案号 31010602000199 医疗器械经营许可证: 沪静药监械经营许20210003号 第二类医疗器械经营备案凭证: 沪静药监械经营备20220042号
营业执照:统一社会信用代码91310108676284138X互联网药品信息服务资格书:(沪)-非经营性-2023-0081
消防排烟风机 华创商务网
美迪医疗网广告业务联系:021-51601230  产品咨询业务联系:021-51601230 传真:021-56532303   美迪医疗网产品咨询 本QQ仅咨询广告和会员业务,不咨询产品和药品等业务美迪医疗网推广业务咨询