DSA 测量技术误差与控制

【 2008-07-25 发布 】 临床报道  

美迪网领先的医疗器械电子商务平台

DSA 测量技术误差与控制
高宗恩 任晓萍 杭鹏 张先军
【摘要】 目的 探讨DSA 测量技术误差的产生与控制。方法 在GE 公司生产的LCV Plus
DSA 机上测得数据。将显示屏分为中央区、中间区及边远区3 个区带, 对不同大小定标在不同检查床
高度及不同点光源与增强器高度( SID) 条件下, 测量标的( 人民币5 角硬币) 的放大、缩小情况。结果
随着标的外移, 由中央区到边远区标的逐渐放大, 且纵向放大比横向放大显著。不同的定标对比,
硬币( 直径20. 4 mm) 和钢球( 直径7. 7 mm) 测量相同标的结果相差较小, 而导管( 4F) 定标有显著的
低估实物倾向。同区带同轴向测量, 测量误差控制在1. 0% ~- 2. 5% 之间。结论 将显示屏划分为
中央区、中间区及边远区将有助于介入医师对测量误差的控制。以定标物的横向做定标来测量标的
较为准确, 同区带同轴向测量误差控制较好。
【关键词】 血管造影术, 数字减影; 放射测量术; 图像处理, 计算机辅助
Assessment of the err or of measur ement t echnique on DSA GAO Zong-en, REN Xiao-ping, HANG
Peng, ZHANG Xian-jun. Imaging center, Shengli Oilfield Central Hospital, Dongying 257034, China
【Abstr act】 Objective To explore the creation and control of measurement technique error on digital
substation angiography ( DSA) . Methods The data was obtained from Advantx LCV Plus DSA system
made by GE Corporation. We divided the screen into three areas, per area account for 1 / 3, ie, central area,
middle area and outlying area. The enlargement rate or reduction rate of the target object was respectively
calculated according to the different calibration, different height of the bed and different X-ray source to
image distance ( SID) . Results The target object was enlarged gradually from the central area to the
outlying area, and the lengthwise enlargement rate was more obvious than transverse. The different of target
object measured by coin ( diameter was 20. 4 mm) with steel ball ( diameter was 7. 7 mm) was not
significance, but the target object was underestimated significantly used the calibration by 4F catheter. When
the target object was measured by the calibration in same area and same axis, the error of measurement
technique was controlled rang from 1. 0% to -2. 5%. Conclusion This systematic investigation suggest that
the screen was divided into the central area, middle area and outlying area will be beneficial to control DSA
measurement error for the interventional physician. The target object was close to real size when it measured
by transverse of the calibration, and the error was better controlled when the calibration was in the same area
and same axis as the target object.
【Key wor ds】 Angiography, digital subtraction; Radiometry; Image processing, computer-assisted

DSA 测量技术误差与控制--普通.rar

网友评论
共计 0 条      
 
我要评论
匿名发表 昵称: Email: 验证码: 点击可刷新
 
    

合作支持:中华医学会 | 中华医院管理学会 | 国家食品药品监督管理家用护理器械商城 | 国药励展展览有限责任公 | 医学装备协会
刊登广告 | 友情链接 | 广告代理商加盟 | 关于美迪 | 法律声明 | 隐私保护 | 网站地图
把美迪网放进收藏夹  把美迪医疗网介绍给我的朋友  给美迪医疗网留言
美迪医疗网广告业务联系:021-51601230 产品咨询业务联系:021-51601230 传真:021-56532303    美迪医疗网业务咨询
互联网药品信息服务许可证:(沪)-经营性-2009-0003   中华人民共和国电信与信息服务业务经营许可证:(沪)B2-20090029 沪ICP备14001091号-8
 
公安备案号 31010602000199 医疗器械经营许可证: 沪静药监械经营许20210003号 第二类医疗器械经营备案凭证: 沪静药监械经营备20220042号
营业执照:统一社会信用代码91310108676284138X互联网药品信息服务资格书:(沪)-非经营性-2023-0081
消防排烟风机 华创商务网
美迪医疗网广告业务联系:021-51601230  产品咨询业务联系:021-51601230 传真:021-56532303   美迪医疗网产品咨询 本QQ仅咨询广告和会员业务,不咨询产品和药品等业务美迪医疗网推广业务咨询